Using Extension Probes with SpanBox5
Published: 2022-06-23 Revised: 2025-05-01
If you need to change your SpanBox5 settings for extension probe use, please follow the steps below:
Edit Settings in Customer Portal
- Log in to the Customer Portal.
- On the Welcome screen, choose Unit Settings.
- Locate the onsite SpanBox by its serial number (SB5-xxx, SBL510/530-xxx) and click the Edit button.
- Select the Multi Products tab, and then the Extension Probe section dropdown.
- Under the Probe Type dropdown, select Extension.
- In the Extension Probe Length field, enter the length of the probe (from the phx42 Quick Connect to the tip filter).
- Click Save.
Confirm Setting Update on SpanBox
Note: The SpanBox must be connected to the network.
- On the onsite SpanBox, select Settings on the home screen menu.
- The Settings response should return a "Success" prompt, confirming the SpanBox is now set for extension probe use.
- Under Daily Cal on the Warm-up screen, refer to the Ext. Probe, and Flow Rate dropdown.
- Calibrate/PreCal/Drift as required.