phx42 – Reset to 15 min. Warm-up Following a Previously-Failed Calibration
Published: 2022-06-28 Revised: 2022-07-01
- Press Daily Calibration.
- On the Cylinders – Active Cylinders screen, tap Manage Cylinders.
- Acknowledge the warning prompt.
- Remain on Active Cylinders for approximately 1 minute.
- Tap Done.
- Resume the calibration process and the phx will go through a 15-minute warm-up.