Can't Get Unit / Product Settings from Customer Portal or Database Cache

Published: 2022-06-23      Revised: 2022-07-01

Two things are happening:

  • The phx42 is not being calibrated on its assigned SpanBox5.
  • The SpanBox5 can't communicate with the LDARtools Customer Portal to download settings.
  1. Check if the SpanBox5 can grab settings from the Support Portal:
    1. Tap Menu.
    2. Tap Settings.
    3. Wait, then do either:
      • If a successful prompt appears, tap OK. Try calibrating again, there may have been a temporary loss of network connection.
      • If you do not get a successful prompt, then proceed to next step.
  2. Verify the SpanBox5 is connected to a wireless network:
    1. First, follow the Manage Wireless Network procedure. Once the wireless network is selected and a Disconnect button appears, the network is connected.
    2. Tap Test Connection to verify the connection is good.
    3. Do either:
      • If successful, navigate to the Home screen, then start Daily Calibration.
      • If unsuccessful, submit a software support ticket and be sure to attach the Cal5.0 logs.