Cal5.0 Not Connected to the SpanBox Base
Published: 2022-06-23 Revised: 2025-11-10
- Verify the green light is illuminated on the power adapter of SpanBox base.
- Verify the power cord is plugged into the SpanBox base.
- Disconnect the power from the SpanBox base, wait 10 seconds, and then reconnect. If the warning cleared, proceed to step 7.
- Verify the USB cable is plugged into the correct USB port on the SpanBox tablet.
- This is mainly for Chuwi tablets because they have two USB-C ports (one for power, one for data). The SpanBox USB needs to be connected to the lowest USB-C port (assuming the serial number SBHU is right-side up).
- Unplug the USB cable from the SpanBox base and tablet.
- Reconnect the USB cable to the tablet and SpanBox base.
- Tap Manual Box Control, if the manual box loads, then toggle on a phx port and listen for a click. If there is no clicking sound, continue to the next step. If the solenoids click, the SpanBox is ready for DailyCal.
- Note: A "Connection Error" may still be present on the Home screen.
- If an error appears when tapping Manual Box Control, continue to the next step.
- Restart Cal5.0 by tapping Restart on the Menu.
- After the restart completes, repeat step 7.