Manually Calibrating the phx21 After a 30-Minute Warm-Up
Published: 2022-07-19 Revised: 2022-07-22
-
Tap Menu.
-
Tap Calibrate.
-
Tap Calibrate at the bottom-left.
-
Type the actual PPM for the cylinder into the PPM field. Zero will be the first gas and is already populated.
-
Apply gas for 10 seconds and then tap Generate.
-
Let the timer count down.
The phx21 will sample the gas, then the PPM value will be added to the list of Current Calibrations on phx21. Repeat Steps 4-6 until all the Calibration Spans have been added, then tap Done.