Probe Testing Adapter Procedure (WI-phx42-115)
Published: 2024-03-18 Revised: 2024-03-18
Click here for a printable version of this procedure.
Items Needed
| Part Name | LDAR Part # |
|---|---|
| phx42 | 4200 |
| Probe Testing Adapter | 4300 |
| phxApp for Self-check | n/a |
Perform the following steps whenever a filter error isn't resolved by following on-screen prompts.
- Remove the probe, attach the probe testing adapter with a clean probe tip filter directly to the phx42, and run a self-check.
- If there are no errors, then the probe was the cause of the issue. Report the probe issue to LDARtools Support.
- If there are errors, submit the issue to LDARtools and comment: "Self-check completed with Probe Testing Adapter."