Skip to main content

Analyzer Drift

Published: 2026-06-18      Revised: 2026-06-18

What This Report Does​

Reports drift check results for analyzers that were drifted using the Standard drift method. A drift check is a spot reading taken during the day to verify that the analyzer is still reading accurately relative to the morning calibration. Each row shows the morning calibration reading for a cylinder, followed by the mid-day (Drift 1) and end-of-day (Drift 2) drift check readings and their percent deviation.

Use this report to verify that analyzers remained within acceptable drift limits throughout the inspection day, and to document drift results for regulatory requirements. This report is in the Calibration folder on the Reports tab.

When to use it

  • Reviewing morning calibration readings alongside intra-day drift results.
  • Confirming that drift checks were performed and that all readings passed.
  • Producing drift documentation for an audit or regulatory submission.
  • Identifying analyzers that exceeded drift limits during a period.

Parameters​

ParameterDescriptionRequiredHow it filters
Start DateStart of the date windowYesCalibration dates on or after this date are included
End DateEnd of the date windowYesCalibration dates on or before this date are included
AnalyzersOne or more analyzers to includeYesOnly records for the selected analyzers appear

Analyzer list is date-filtered. The Analyzers parameter only shows analyzers with a Standard drift record in the database for the selected date range.

Columns​

ColumnWhat it showsFormat / Notes
AnalyzerAnalyzer device nameText
DateDate the calibration and drift checks were performedDate; facility local time
Tech IdTechnician who performed the calibrationText
CalibrateTime of the morning calibration readingTime (H:MM:SS AM/PM)
CylinderCylinder label or serial numberText
Cyl. ExpCylinder expiration dateDate
Conc.Cylinder gas concentration in PPMInteger
ReadingMorning calibration reading in PPMDecimal
PassWhether the morning calibration passed"Yes" or "No"
Drift1Time of the mid-day drift check (Drift 1)Time (H:MM:SS AM/PM); blank if no mid-day drift performed
Drift1 ResultsMid-day drift check reading in PPMDecimal; blank if no mid-day drift
Pct.Mid-day drift percent deviationPercentage; blank if no mid-day drift
PassWhether the mid-day drift passed"Yes" or "No"; blank if no mid-day drift
Drift2Time of the end-of-day drift check (Drift 2)Time (H:MM:SS AM/PM); blank if no end-of-day drift
Drift2 ResultsEnd-of-day drift check reading in PPMDecimal; blank if no end-of-day drift
Pct.End-of-day drift percent deviationPercentage; blank if no end-of-day drift
PassWhether the end-of-day drift passed"Yes" or "No"; blank if no end-of-day drift
Ext ProbeWhether an extension probe was used"Yes" or "No"

What's Included and Excluded​

Included:

  • Calibration records (morning readings) for the selected analyzers within the date range.
  • Drift 1 (mid-day) readings matched to the morning calibration.
  • Drift 2 (end-of-day) readings matched to the morning calibration.

Excluded:

  • Analyzers not selected.
  • Records outside the date range.
  • Analyzers drifted using the VVa method — a three-reading EPA drift method (40 CFR Part 60, Appendix F, Method VVa) — those appear in the VVa Drift report.
  • Zero gas rows do not show drift percent values (no drift calculation for zero gas).

How Key Values are Calculated​

Drift Percent

The percent deviation between the drift reading and the morning calibration is calculated according to the drift method configured in the analyzer's calibration settings:

SettingFormula
VVa 60.485a(Drift − Calibration Reading) ÷ Cylinder Concentration × 100
Calibration PPM(Drift − Calibration Reading) ÷ Calibration Reading × 100
Cylinder PPM (default)(Drift − Cylinder Concentration) ÷ Cylinder Concentration × 100

A value of 0% means the drift reading exactly matched the reference value. Negative values indicate the analyzer is reading lower than expected; positive values indicate higher.

Drift 1 vs Drift 2

  • Drift 1 is the mid-day (Noon) drift check, taken partway through the inspection shift.
  • Drift 2 is the end-of-day drift check, taken at the end of the shift.

Not all sessions will have both; either drift may be blank if that check was not performed.

Tips and Common Questions​

  • "Drift columns are blank for some rows." If no mid-day or end-of-day drift was performed for that calibration session, the corresponding Drift columns will be blank. Zero gas rows also never show drift percent values.
  • "There are multiple rows per analyzer per date." Each cylinder used in the morning calibration produces one row. A typical session includes a zero gas and several span gases.
  • "Ext Probe = Yes but the morning calibration used Standard." Ext Probe reflects whether the calibration activity used an extension probe. A session can include both Standard and Extension probe readings on different gas cylinders.
  • "The Pct. sign is negative — does that mean it failed?" A negative percent deviation means the analyzer is reading lower than expected. Whether this is a pass or fail depends on your site's drift limit (typically ±10% or ±20% depending on the regulation). Check the Pass column for the actual pass/fail result.
  • "This report only shows Standard drift — where is VVa drift?" Analyzers drifted using the VVa method appear in the VVa Drift or VVa Drift – Form 2 reports instead.

Sample Output​

Report run for Example Facility, Start Date: 9/22/2022, End Date: 9/22/2022, Analyzers: phx42-3536, phx42-3584, phx42-3590, phx42-3599, and phx42-5389.

Analyzer Drift sample output

What the rows illustrate:

  • CT-24077 (zero gas) — Zero gas row; no drift percent shown. Drift1/Drift2 columns are blank.
  • phx42-3536 (CT8327) — Drift 1 at 1:41 PM showing −3.98%; Drift 2 at 4:36 PM showing −0.31%. Both pass. The analyzer is drifting lower throughout the day but remains within limits.
  • phx42-3536 (T7790071) — Smaller drift values than before across both mid-day and end-of-day checks; well within acceptable limits.

Output Format​

One row per gas cylinder per morning calibration session. Drift 1 and Drift 2 readings appear as additional columns in the same row. Suitable for on-screen viewing, PDF, and Excel export.

  • VVa Drift — drift checks performed using the VVa method (three readings per drift event, NSPS (New Source Performance Standards) VVa and Consent Decree compliance columns).
  • VVa Drift – Form 2 — simplified VVa drift format with Cal Note column.
  • Daily Calibration — the morning calibration readings without the drift check columns.